Bending Fatigue Tests of Osteosynthesis Devices in Pseudo Physiological Solution

In-page menu

Evaluation of durability of bone plates in simulated body fluid.

Outline of Measurement Technology

The durability of osteosynthesis parts is evaluated by 4-point bending or 3-point bending in water or simulated body fluid (SBF). Test jigs and conditions are designed referring to JIS T 0312 or ASTM F382, which are test standards for osteosynthesis parts in the atmosphere, and tests are performed in SBF kept at the temperature of 37℃.

Outline of Test

Test environment 37 ± 2℃, in liquid (distilled water or simulated body fluid)
Test method Bending fatigue test (3-point bending or 4-point bending)
Max. load 10kN
Max. cycle frequency 2Hz *
Types of test specimens Bone plates, CHS (compression hip screw), screws, lag screws

*For other test frequencies, please consult JFE-TEC.

Example of Measurement of Bending Fatigue Strength of Intramedullary Nail

Example of Test

Cycle frequency : 1Hz
Load ratio : 0.1
Test environment : 37°C Phosphate buffered saline (PBS)
Schematic diagram of fatigue test
Schematic diagram of fatigue test

We can also prepare test reports in English and a number of other languages!

In addition to Japanese, we can also provide test reports in accordance with ASTM, ISO and other standards in English.

We can also provide expert translations in many other languages. Please discuss your language needs with us when making an inquiry or request.