Evaluation of Medical Equipment/Implant Materials
Measurement of Static Strength and Fatigue Strength of Interconnection Mechanism of Spinal Rods
Measurement of interconnection strength of spinal implants and loosening resistance of parts
Outline of Static Strength and Fatigue Strength Measurement Technology
Spinal connection parts (hooks, screws, bands, etc.) and vertical elements (rods, plates) are combined in accordance with the manufacturer's instructions, and static tests and fatigue tests of the spinal implant structure are performed.
Connection Strength Measurement Conditions
Corresponding standard | ASTM F1798 |
---|---|
Test environment | Constant temperature, constant humidity, in atmosphere, in saline (37 ± 2)°C |
Test methods | Tensile test, compression test, bending test, torsion test |
Max. load | 10 kN, 10 N-m |
Max. cycle frequency | 16Hz |
Max. number of cycles | 2.5 x 106 |
Connection Strength Measurement Methods
We can also prepare test reports in English and a number of other languages!
In addition to Japanese, we can also provide test reports in accordance with ASTM, ISO and other standards in English.
We can also provide expert translations in many other languages. Please discuss your language needs with us when making an inquiry or request.